PrecisionBind Human Colony Stimulating Factor 2 Granulocyte Macrophage (GM-CSF) ELISA (NIBSC/WHO calibrated)

SKU: KB1048

Rs 38,000.00

Enzyme Immunoassay for the estimation of Colony Stimulating Factor 2 Granulocyte Macrophage (GM-CSF) in human serum, plasma, cerebrospinal fluid and other biological samples.

PrecisionBind ELISA are matrix-evaluated cytokine immunoassays designed for immune monitoring and translational research, with reference-standard calibration (NIBSC/WHO where available) and performance documentation aligned to applied bioanalysis (ICH M12 for linearity, recovery, precision, and matrix effects). RUO only.

Sample matrices Validated: Serum, Plasma, Cerebrospinal Fluid, Cell Culture Supernatant and Cell / Tissue Lysates

Assay Range: 15.625 – 1000 pg/ml | Analytical Validation: LOD: ~3.03 pg/ml/ LOQ: 9.17 pg/ml / IC50: ~439.7 pg/ml

Availability: (within US) 1 – 2 Days, (outside US) 3 – 4 weeks

Pack Size: 1 x 96 wells, break-apart

5 in stock

 

Guaranteed Safe Checkout

Accepted payment methods

All our kits are manufactured under ISO 13485.

The precisionbind Human Colony Stimulating Factor 2 Granulocyte Macrophage (GM-CSF) ELISA includes features like:
– Ready to use protocol with break-apart wells for ease of use
– Standardisation and High Reproducibility
– Lot to Lot Consistency
– Accuracy and Precision

Principle Of Assay: This ELISA is a sandwich immunoassay. Antibodies are coated on 96 well plates. The antigen protein present in sample and standard respectively bind to the coated wells. The wells are washed and an antibody:HRP Conjugate is added which binds to the bound complex in the well. Washing is performed to remove any unbound material. TMB substrate is added and the enzyme reaction is stopped by dispensing of stop solution into the wells. The optical density (OD) of the solution at 450 nm is directly proportional to the amount of antigen protein present in the standard or samples.

Disclaimer: The data indicated herein with specifications are changed from time to time at time of production of the assay. We request you to confirm the specifications including the assay range and procedure as per the most current IFU (instructions for use) accompanying the assay kit.

Validated against seven points for a “GOLD RING” Standard Quality ELISA – the benchmark sign for Krishgen Quality. The The precisionbind ELISA kits are used for assessing the specific biomarker in samples analytes which may be human serum, plasma, biological fluids and cell culture supernatant.

The kit uses indirect sandwich assay with double antibodies – capture and detection to ensure a high degree of sensitivity and specificity in the estimation of Colony Stimulating Factor 2 Granulocyte Macrophage (GM-CSF);

About PrecisionBind:
PrecisionBind is Krishgen`s cytokine ELISA platform developed for bioanalytical robustness in translational research, including immuno-oncology and cell & gene therapy programs. Where applicable, kit standards are calibrated against NIBSC/WHO international standards to support unit traceability and inter-study comparability.

Assays are characterized following ligand-binding assay performance expectations, including:

  • Defined assay range and sensitivity (LOD/LOQ where reported)
  • Intra- and inter-assay precision
  • Cross-matrix performance characterization (serum, plasma, CSF where applicable), including parallelism and recovery assessment, to support quantitative comparability across biological matrices.All kits are Research Use Only (RUO) and not intended for diagnostic use.

Validation Parameters and Acceptance Criteria

1. GM-CSF Cmax Values and Recommended ELISA Range
This table summarizes GM-CSF Cmax levels across diseases and suggests corresponding ELISA
working ranges.

Note: Assay sensitivity < 2 pg/mL recommended for baseline detection; upper limit ≥ 2,000 pg/ml
advised for CRS monitoring.

2. Specificity and Selectivity

2.1 Specificity
The capture antibody and detection antibody are both specific to GM-CSF and are monoclonal
antibodies. They show a high affinity to bind to native as well as recombinant GM-CSF.

2.2 Selectivity
The ELISA has no or low cross reactivity to IL‑1β (IL-1beta), IL‑6, or TNF‑β (TNFbeta).

2.3 NIBSC validation
The standard used in the kit is calibrated against an international standard from the National
Institute of Biological Standards and Control (NIBSC), Potters Bar, Hertfordshire EN6 3QG, UK. 1
ng of supplied standard equals 6 U of 88/646 NIBSC-standard.

Therefore 1000 pg/ml is equivalent to 6 U of GM-CSF as per NIBSC.

2.4 LOD, LOQ and IC50
LOD (Limit of Detection)
The lowest analyte concentration that can be reliably distinguished from blank/background noise
but not necessarily quantified precisely.

Statistically:
LOD = Mean of Blank + 3X SD of Blank
(3σ criterion is most common).
LOD for PrecisionBind Human GM-CSF ELISA = 3.03 pg/ml

LOQ (Limit of Quantitation)
The lowest analyte concentration that can be quantified with acceptable accuracy and precision.

Statistically:
LOQ = Mean of Blank + 10X SD of Blank
(10σ criterion is most common).

LOQ for PrecisionBind Human GM-CSF ELISA – 9.17 pg/ml

IC50 in ELISA (Half Maximal Inhibitory Concentration)
IC50 = The concentration of an inhibitor (drug, antibody, compound) required to reduce the signal
(e.g., binding, enzymatic activity) by 50% compared to the maximum signal in the assay.

In ELISA, this is commonly used for:
Neutralization ELISA: Quantifies potency of antibodies inhibiting target–ligand interaction.
Drug Potency Testing: Measures concentration at which drug inhibits 50% of target activity.

IC50 for PrecisionBind Human GM-CSF ELISA = ~439.7 pg/ml

Regulatory Note:
LOD *S/N ≥ 3:1, LOQ ≥ 10:1, %CV ≤ 20%      *S/N = Signal / Noise Ratio

3. Linearity and Range

 

4. Accuracy and Precision (Intra / Inter-Assay)

A) Intra-Assay:

B) Inter Assay:

5. Parallelism and Matrix Effect
Sample Dilution factor – Human Serum, Human Plasma and Human CSF samples have been
tested. Neat samples can be run directly.
Neat Human Serum, Human Plasma and Human CSF were spiked with 500 pg/ml Human GMCSF and ELISA assay was run.

A) Serum:

B) Plasma:

C) Cerebrospinal Fluid (CSF):

Results:
i) Parallelism is maintained across the 1:1 to 1:16 dilutions.
ii) % Recovery for most dilutions falls within the acceptable range of 80%–120%.
iii) No significant matrix effect observed at higher dilutions.
iv) The PrecisionBind Human GM-CSF ELISA kit was tested for matrix effect on human serum,
plasma, CSF and physiological buffer 7.4 to mimic tear fluid samples.

6. Sample Handling and Storage Conditions

A.) Sample collection and handling:
Specimens should be clear and non-hemolyzed. Samples should be run at a number of dilutions
to ensure accurate quantitation.

Cell Culture Supernatant: If necessary, centrifuge to remove debris prior to analysis. Samples
can be stored at temperature <-20°C. Avoid repeated freeze/thaw cycles.

Serum: Use a serum separator tube and allow clotting for 30 minutes, then centrifuge for 10
minutes at 1000 x g. Remove serum layer and assay immediately or store serum samples at
temperature <-20°C. Avoid repeated freeze/thaw cycles.

Plasma: Collect blood sample in a citrate, heparin or EDTA containing tube. Centrifuge for 10
minutes at 1000 x g within 30 minutes of collection. Assay immediately or store plasma samples at
temperature <-20°C. Avoid repeated freeze/thaw cycles.

B) Storage conditions:
Store main kit components at 2-8°C.
Store recombinant lyophilized standard at 2-8°C. Upon reconstitution aliquot standards into
polypropylene vials and store at -20°C as per assay requirements. Do not freeze thaw for more
than two times.

Before using, bring all components to room temperature (18-25°C). Upon assay completion return
all components to appropriate storage conditions.

Graphs, Maps and Appendices:

 

Matrix Effect Heat Map

 

Determined Limits for Acceptance according to EMA/FDA and CLSI regulations

References
FDA Guidance for Industry: Immunogenicity Assessment for Therapeutic Protein Products (2014).
EMA Guideline on Biosimilar Monoclonal Antibodies (2012)

Additional information

Pack Type

1 x 96 wells, break-apart

Species Reactivity

Human

Sample Type

Serum, Plasma, Cerebrospinal Fluid, synthetic Tear Fluid, Cell Culture Supernatant and other biological samples. Validated for citrated/EDTA Plasma only. Check with us if you wish to have the validation data for any other specific matrices.

Assay Range

15.625 – 1000 pg/ml

Standards Available

0; 15.625; 31.25; 62.5; 125; 250; 500; 1000 pg/ml

Analytical Validation

LOD: ~3.03 pg/ml/ LOQ: 9.17 pg/ml / IC50: ~439.7 pg/ml

Principle Of Assay

Sandwich ELISA, Double Antibody

Detection Method

Absorbance measured at 450nm, Colorimetric

Conjugate-Enzyme Reaction

and dynamic range, balancing speed, sensitivity

Regulatory Status

For Research Use Only (RUO).

Shipping Temperature

Shipped at Room Temperature or in gel pack to maintain temperature at 2-8 degrees Celsius. Some components may be shipped to -20 degrees Celsius

Storage Temperature

if all the wells in the kit are not being used, store the unused wells in the foil pouch containing the desiccant pack, The kit (2-8 Degree Celsius) and its components are to be stored as indicated in the IFU (instructions for use). To ensure quality of your results

ELISA Interpretation

4PL (2nd order) or Cubic Spline Graph. Software (MS Excel) available, on request, Quantitative Results Intepretation, to support interpretation of results.

Shelf Life Available

8 -10 months at time of shipping

Alternate Names / Synonyms

Granulocyte-macrophage colony-stimulating factor; GM-CSF; Molgramostin; Sargramostim; CSF2; GMCSF

Research Area

Immunology, Signal Transduction, Cardiovascular; Cell Biology; Immunology; Gene Therapy, Cell Therapy

Disclaimer

The data indicated herein with specifications are changed from time to time at time of production. Please refer the most current IFU (instructions for use) accompanying the kit.

Documents

PrecisionBind Human GM-CSF-ELISA-Validation_Guide_2025_ver 1.0